SP-950 is a syringe pump for medical use, and it adopts microcomputer system and microelectronic technology, features in high-intelligence and high-safety. LCD can display each parameter in real-time, with multi-function alarm, and alarm with visual, auditory and character prompt. It takes the advantages of easy operation, steady work, high accuracy and checking the syringe specification automatically. And now, it has been widely applied in each vein injection in hospitals for clinical nursing, such as ICU, CCU, neonatology, obstetrics and gynecology, pediatrics, internal medicine, surgery, operating room, emergency room.
CONTEC SP-950
10ml/20ml/30ml/50ml (CNS)
10ml - 0.1ml/h-100ml/h, 20ml - 0.1ml/h-200ml/h, 30ml - 0.1ml/h-300ml/h, 50ml - 0.1ml/h-500ml/h.
10ml - 100ml/h, 20ml - 200ml/h, 30ml - 300ml/h, 50ml - 500ml/h.
0ml-999.9ml.
±3%.
Three grades are adjustable - Low: 40KPa ± 16KPa, Middle: 70KPa ± 30KPa, High: 100KPa± 35KPa.
0.1ml/h~5ml/h, adjustable
Class Ⅱ equipment, internally powered type BF.
At non-operated state, Near empty, Empty, enter KVO mode, Syringe fall-off, Occlusion, Pressure failure, Motor failure, Low battery, Communication abnormal.
20VA~30VA, the device can continuously work for at least two hours with the injection rate of 5ml/h or 500ml/h after fully charged.
AC100~240V 50/60Hz.
Rechargeable lithium battery, 7.4V, 3500mAh.
Temperature +5~40°C; Atmospheric pressure 860hPa~1060h; Relative humidity 20%~90%; no strong vibration and no corrosive gas.
Temperature: -20°C~+55°C; Relative humidity ≤90%, no condensation; No corrosive gas and well-ventilated.
310mm*140mm*135mm(L×W×H).
1.7kg.
A bracket
A power cord
A user manual
Two fuses